Hanna Instruments HI 8427 User Manual
Page 3

5
4
FUNCTIONAL DESCRIPTION HI 931001
Range
pH
0.00 to 14.00
mV
-1000 to 1000
Resolution
pH
0.01
mV
1
Accuracy
pH
±0.01
(@20°C/68°F) mV
±1
EMC Typical pH
±0.02
Deviation
mV
±2
Temperature
All output values simulated
Compensation
at 25°C (77°F)
Battery Type
9V
Life
500 hours of continuous use
Environment
0 to 50°C (32 to 122°F); 95% RH
Dimensions
180 x 83 x 40 mm (7.1 x 3.3 x 1.6")
Weight
320 g (11.3 oz.)
HI 8427
pH/mV CALIBRATOR
LOW BATTERY
LEAKAGE
pH2
pH4
pH7
pH10
pH12
pH14
1900
mV
1900
mV
350
mV
-350
mV
OFF
pH0
HIGH
IMPEDANCE
°C
0
10
20
30
40
50
9
7
8
3
1
2
5
3
4
6
Range
pH
0, 2, 4, 7, 10, 12, 14
mV
-1900, -350, 350, 1900
Accuracy
pH
±0.1
(@20°C/68°F)
mV
±5
Temperature
Manual from 0 to 50°C
Compensation
(32 to 122°F)
High Impedance Test
10
9
Ω
Battery Type
9V
Life
100 hours of continuous use
Environment
0 to 50°C (32 to 122°F); 95% RH
Dimensions
180x83x40 mm (7.1 x 3.3 x 1.6")
Weight
255 g (9 oz.)
FUNCTIONAL DESCRIPTION HI 8427
1. Rotary Knob
2. Switch Off Selection
3. pH Outputs Selection
4. mV Outputs Selection
5. Temperature Compensation Knob
6. High Impedance Key
7. Low Battery LED
8. Leakage LED
9. BNC Socket
1. LCD Display
2. ON/OFF Key
3. Range Selection Key
4. pH/mV Coarse Setting Knob
5. pH/mV Fine Setting Knob
6. BNC Socket