Hanna Instruments HI 98230 User Manual
Page 6

11
10
code 40: Auto OFF
code 41: Battery test
code 50: Baud rate
code 60: Release code
code 70: Celsius or Fahrenheit
code 99: Password
Some of the most important functions are explained below in a step
by step sequence.
TO SCAN LOGGED DATA
CODE 00 - Lot Data Printing / Scrolling
• Select the code 00.
• The message "Lot data Printing" will scroll twice on LCD.
• The upper LCD will then display L 00
with the 00 blinking.
• Set the desired lot by the arrow keys. L00 is the lot of data of the
"log on demand" and L01 to L16 are the lots of the "timed log".
• Press the ALT and CFM keys to confirm the lot number.
• If the lot doesn't contain data, the "no data" message will scroll
across the LCD twice and the meter will return to setup mode.
• If the lot contains one or more data the
LCD will display the sample number in its
upper part and Sn in the lower part.
Note: In the L00 lot (log on demand) the sample number will be
displayed with 3 digits (001).
• Select the sample number to scan by the arrow keys.
Printing Logged Data
• Press ALT and CFM to print logged data.
• If the selected sample number is invalid (equal to 0 or bigger
than the number of samples), the "Err" message will be displayed
for a few seconds.
• If the sample number is correct, the samples starting from the
selected one to the last sample of the lot will be printed. To stop
printing before the last sample is reached, press and hold the ALT
and PAPER keys until the printer stops.
• During printout the LCD will display the sample number being
Code
Valid values
Default
10 Show pH GLP data
11 Show ORP GLP data (HI 98240 only)
12 Print GLP data
13 Calibration alarm time-out
00 to 99 days
00
00 means option disabled
14 Instrument ID code
0000 to 9999
0000
30 Current time
1
hh:mm
00:00
31 Current day
1
dd
01
32 Current month
1
MM
01
33 Current year
1
YYYY
1998
40 Auto-Off/Power down time-out 5,10,15,30,45,60
5
41 Battery level test
50 RS232 baud rate
1200, 2400, 4800
4800
60 Firmware version
70 Celsius/Fahrenheit selection
°C or °F
°C
99 Password
2
0000 to 9999
0000
Note: If a wrong code is entered the "Err" message is displayed on
LCD for a few seconds.
1
The meter automatically checks for entered time/date accuracy as follows:
0
≤
hh
≤
23; 0
≤
mm
≤
59; 01
≤
dd
≤
28/29/30/31; 1
≤
MM
≤
12; 1998
≤
YYYY
≤
2097.
2
To change the password, the correct code must be entered first. If the password
has been forgotten, the password protected features are no longer accessible;
in this case contact your nearest Hanna Service Center.
SETUP MESSAGES LIST
code. 00: Lot data Printing
code 01: Lot table Printing
code 02: Printer enable
code 03: Log Interval
code 05: Press "ALT CFM" to delete Lot 00 or "ALT FNC" to escape
code 06: Press "ALT CFM" to delete Lot 01-16 or "ALT FNC" to escape
code 10: pH GLP
code 11: Volt GLP (HI 98240 only)
code 12: GLP data printing
code 13: Calibration alarm time-out
code 14: Instrument ID Code
code 30: Hour - Minute
code 31: Day
code 32: Month
code 33: Year