beautypg.com

Hanna Instruments pH 20 User Manual

Page 4

background image

7

6

Note: If measurements are taken in different samples successively, it is

recommended to clean the electrode thoroughly to avoid cross-
contamination. After cleaning, rinse the electrode with water
and then with some of the solution to be tested.

MANUAL TEMPERATURE COMPENSATION

• When the temperature probe is not connected, the default

temperature value is 25°C.

• To temperature compensate the pH read-

ing, measure the temperature of the solution
with a

ChecktempC or another accurate

thermometer. E.g. 20°C.

• Press the RANGE key to enter the °C mode and adjust the

displayed value by using the arrow keys.

TAKING ORP MEASUREMENTS (for pH21 only)

• The meter is factory calibrated for ORP (mV) range. Please

contact the nearest Hanna Service Center for recalibration,
if needed.

• Connect an ORP electrode (optional, see "Accessories" section) to the

BNC connector on the rear panel (see page 4).

• Connect the 12VDC power adapter to the instrument (see page 4)

and then to the mains.

• Press the ON/OFF key to turn the meter

on.

• Press the RANGE key to select the ORP

(mV) mode.

• Remove the electrode protective cap, then immerse the ORP

electrode in the sample to be tested.

Note: The electrode should be submerged approximately 4 cm (1½")

into the solution.

Note: With refillable electrodes, to obtain a faster response unscrew

the refill hole cap during measurements.

• Shake briefly and wait a few seconds for

the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the ORP (mV) value.

TAKING pH MEASUREMENTS

• Make sure that the instrument has been calibrated for pH before

taking measurements (see "pH Calibration" section for details).

• Connect the supplied HI 1110B pH electrode to the BNC connec-

tor on the rear panel (see page 4).

• If automatic temperature compensation is required, connect the

HI 7662 temperature probe (optional) to the proper socket on
the rear panel (see page 4).

• Connect the 12VDC power adapter to the instrument (see page 4)

and then to the mains.

• Press the ON/OFF key to turn the meter

on.

• Press the RANGE key to select the pH

mode, if necessary.

• Remove the electrode protective cap, then immerse the pH

electrode and the temperature probe in the sample to be tested.

Note: If the pH electrode is dry, immerse it in HI 70300 storage

solution for at least one hour to reactivate it.

Note: The pH electrode should be submerged

approximately 4 cm (1½") into the
solution and the temperature probe (if
used) should be located as close as
possible to the electrode.

Note: With refillable electrodes, to obtain a faster response unscrew

the refill hole cap during measurements.

• Shake briefly and wait a few seconds for

the reading to stabilize, i.e. when the
hourglass symbol stops blinking. The LCD
will show the pH value automatically com-
pensated for temperature variations.

• After measurements, turn the meter off by pressing the ON/OFF

key, disconnect the power adapter from the mains and the pH
electrode from the instrument.

• Store the electrode with a few drops of HI 70300 solution in the

protective cap.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STORING
PURPOSES.

OPERATIONAL GUIDE

This manual is related to the following products: