beautypg.com

Mportant – EXFO FTB-2 User Manual

Page 3

background image

P/N: 1066920

July 2014

3/6

To install a power meter:
1.
Turn off your unit.

2. Disconnect the unit from its external power source and remove the

batteries.

3. Put on a wrist (or an ankle) antistatic strap and ensure that the end of

its wire is grounded properly.

4. Place the unit vertically so that its right panel is facing you, and then

locate the protective cover that hides the power meter bay.

5. Using a screwdriver, remove the two screws of the protective cover.

Note: Keep the screws close at hand because you will need them to secure the

power meter in place later.

6. Remove the protective cover.

I

MPORTANT

Do not discard the protective cover. You will need it to protect the
power meter bay if you ever send the power meter for
recalibration.

Screws holding the

protective cover in place