Technical Specifications
33
Model FOT-302
FOT-302X
Power meter port
b
Ge
GeX
Power range (dBm)
c
10 to —60
26 to —50
Range displayed (dBm)
Down to —65
Down to —50
Number of calibrated wavelengths
d
10
10
Power uncertainty
e
± 5 % ± 1 nW
± 5 % ± 10 nW
Resolution (dB)
0.01
f
0.01
g
Automatic offset nulling
h
Yes
Yes
Warmup time (s)
h, i
0
0
Display units
dB/dBm/W
dB/dBm/W
Automatic wavelength recognition
j
Yes
Yes
Screen refresh rate (Hz)
3
3
Tone detection (Hz)
270, 1 k, 2 k
270, 1 k, 2 k
Battery life (hours) (typical)
260
260
Warranty and recommended calibration interval (years) 3
3
Model
k
23BL
12D
235BL
Central wavelength (nm)
1310 ± 20
850 ± 25/
1310 ± 20
1550 ± 20
1300 + 50/—10
1490 ± 10
1550 ± 20
Spectral width (nm)
l
) 5
50/135
) 5
Output power (dBm)
* 1/* 1
* —20/* —20 (62.5/125 +m)
* 1/* —4.5/*—3
Power stability (dB)
m
8 hours
± 0.1
± 0.1
± 0.1
Battery life (hours)
n
120
120
120
Automatic wavelength recognition
Yes
Yes
Yes
Tone generation (Hz)
270, 1 k, 2 k
270, 1 k, 2 k
270, 1 k, 2 k
Warranty and recommended calibration interval (years) 3
3
3
SPECIFICATIONS
a
GENERAL SPECIFICATIONS
Size (H x W x D)
185 mm x 100 mm x 55 mm (7
1
/
4
in x 4 in x 2
1
/
8
in)
Weight
0.4
kg
(0.9
lb)
Temperature operating —10 °C to 50 °C
(14 °F to 122 °F)
storage
—40 °C to 70 °C
(—40 °F to 158 °F)
Relative humidity
0 % to 95 % non-condensing
STANDARD ACCESSORIES
User guide, Certificate of Calibration, instrument stickers in six languages,
AC adapter, EUI-XX, connector adapter (FOA-XX), three AA batteries, wrist
strap, alcohol cleaning pads.
SAFETY
21 CFR 1040.10 and IEC 60825-1:1993+A1:1997+A2:2001:
CLASS 1 LASER PRODUCT CLASS 1 LED PRODUCT
NOTES
a. Guaranteed
unless
otherwise
specified.
b.
All specifications valid at 1550 nm and 23 °C ± 1 °C, with
an FC connector.
c. In CW mode; sensitivity defined as 6 x rms noise level.
d.
Wavelengths: 830 nm, 850 nm, 980 nm, 1300 nm, 1310
nm, 1450 nm, 1490 nm,
1550 nm, 1590 nm and 1625 nm.
e. Traceable to national standards; FOT-302X: up to 20 dBm.
f.
From 10 dBm to
—
50 dBm.
g. From 26 dBm to
—
35 dBm.
h.
Power of > —40 dBm for FOT-302, and of >
—
25 dBm for
FOT-302X.
i.
For ± 0.05 dB, from 18 °C to 28 °C.
j.
At 850 nm, 1300 nm, 1310 nm, 1490 nm, 1550 nm and
1625 nm; power > —50 dBm for FOT-302, and
> —40 dBm (typical) for FOT-302X.
k.
All specifications valid at 23 °C ± 1 °C, with an FC
connector.
l.
rms for lasers and —3 dB width for LEDs; typical values for
LEDs.
m.
After 15 minutes warmup; expressed as ± half the difference
between the maximum and minimum values measured during
the period, with an APC connector on the power meter.
n. Typical
autonomy
in
Auto
mode.