beautypg.com

EXFO DC0091A User Manual

Page 3

background image

DC0091A

November 2001

3/3

Note: If all the solutions mentioned above do not seem to work, make sure the

FTB-100/FTB-100B is turned off and cut all power sources (unplug the
power supply and remove the 9 V and main batteries) for 10 seconds.
Then, reinsert the batteries and power up the unit. If the problem persist,
call EXFO.

The 9 V battery must

be replaced regularly.

See. Battery Maintenance
Recommendations
on page 1

When pressing the

PUSH button on the

battery, the LEDs do

not light up.

The battery voltage is

below 5.8 V and the

NiMH main battery is

uncalibrated.

Perform a complete battery

recalibration

b

.

The performance of

the unit degrades

noticeably

OR

The “Storage capacity
is low and may slow
down the operation of
the unit. Please delete
unnecessary
files
”message is

displayed on-screen

upon startup.

The internal flash ROM

has less than 512 KB of

free memory left.

Free up more memory in the

FTB-100/FTB-100B’s internal flash

ROM by deleting some stored trace

files or by transferring them to a PC

Card or floppy disk.

a.

The screen is only completely dark on the color LCD. On the monochrome LCD, you can always

distinguish the screen, even when the backlight is completely off.

b.

A complete battery recalibration consists of a full discharge, immediately followed by a full charge, then

followed by a full discharge.

Problem

Cause

Solution