3.0 Installation

3.1 Mounting of Instrument

The M

ED

A

IR

2000 should be located near the pipe or tank containing the air to be monitored, and upstream from

where the air is being used. So that, the air sample enters the instrument before it reaches the users.
Upright (plumb and level) vertical orientation of the instrument is necessary for proper operation. Mount the
instrument on an appropriate vertical surface using the mounting flanges provided. Avoid areas with excessive
vibration. The holes in the flanges are 0.31 inch in diameter and form a 6 x 10.75 inch rectangle. See Figure 3.

ALARM
ACKNOWLEDGE

AUDIO
DEFEAT

FAULT

POWER

SELECT

OPTION

Dimensions are in inches.

Figure 3: M

ED

A

IR

2000

Mounting Dimensions

Mounting Holes
0.31” dia. 4 places