7.4 Hydrocarbon Filter Replacement

The charcoal absorption filter for hydrocarbons will need periodic replacement. The length of time between
replacement depends on the instrument usage and VOC (Volatile Organic Compounds) level in the breathing air supply.

For ISA-RAL monitors that are in continuos air streams, 24 hours, 7 days a week 365 days a year it is recommended
that the filter be replaced when any of these conditions occur:

At least annually

Any time fluid is found inside the filter

Any time the filter is damaged

If any odor is detected in the breathing mask

8.0 Specifications

8.0 Specifications

Specifications are subject change without notice

M

AXIMUM

P

OWER

117 Vac, 60 Hz, 25 watts

12 Vdc, 25 watts

R

ELAY

C

URRENT

2 amp steady, 5 amp surge @ 110Vac

MOS S

ENSOR

L

IFE

Up to 3 years in clean air (no oil)

R

ESPONSE

T

IME

Up to 5 minutes for low-level carbon monoxide gas
calibrations

M

AXIMUM

A

IR

L

INE

P

RESSURE

250

PSIG

(ISA-44 RAL-OD)

S

AMPLE

F

LOW

R

ATE

Approximately .4

SCFH

(

Standard Cubic Foot Hour)

O

XYGEN

M

ICRO

-F

UEL

C

ELL

linearity error (0 – 100 % oxygen)

less than 0.5%

at 0% oxygen

less than 2.0 ua

nominal (in air @ 25 degrees C)

175 ua

O

UTPUT

variation from nominal

+20%

R

ESPONSE

T

IME

@ 25 D

EGREES

C (T

YPICAL

) 90%

30 seconds

O

PERATING

T

EMPERATURE

R

ANGE

32 - 125 degrees F

E

XPECTED

L

IFE

up to 12 - 14 months in air

ENMET W

ARRANTY

6 months

8.1 Typical Calibrations

Hazardous Gas

Low Level Alarm Threshold

High Level Alarm Threshold

Carbon Monoxide

10 or 20 ppm

20 or 50 ppm

Oxygen Deficiency

19.5%

8.2 Options and Variations

O

PTIONS

220 Vac power input connection

V

ARIATIONS

a)

Available in an explosion-proof instrument housing, Models ISA-44-RALE-OD and ISA-44AHE-OD. With this, the
equipment is suitable for operation in Class I, Division I, Group C and D atmosphere, as defined in the National
Electrical Code. Units for Group B are available upon special order. DO NOT use the standard ISA-44RAL-OD
where you are required by the National Electrical Code or local codes to use an explosion-proof variety.

b)

ENMET Filtration Panels filter oil, water and solid particles from compressed air.

AFS-50:

equipped to distribute air to four respirators.

Ask for product bulletin on air purification and monitoring systems.