ENMET ISA-40 User Manual

Page 7

background image

ISA-40 Oxygen Monitor

ENMET Corporation

5

ISA-40 Interior View

Figure 3: Oxygen Cell Installation

3.2 Oxygen Cell Installation

For some ISA-40 systems, single channel systems, and all multichannel systems, remote sensors are connected to the
control unit via wiring. To connect these assemblies to the control unit, see the Wiring Information Tables 2 & 2A
Section 3.2.2. The transducer cells are each packaged in a vapor proof plastic bag to help prevent depletion of useful
life before installation.

3.2.1 Oxygen Cell Installation Inside of Control Unit

The oxygen detection transducer cell is mounted inside the unit. The cell is packaged in a vapor proof plastic bag to
help prevent depletion of useful life before installation. Do not open or puncture the bag before installation to insure
best performance of cell.

1. Remove 4 cover screws to open the control unit and remove screws to remove cell circuit board. See Figure 1 & 3.

2. Carefully remove the cell from the package.

3. Remove the metal shorting clip (or spring) from cell pins. See Figure 3.

4. Plug the cell into the component side of the cell circuit board; center pin to center hole. See Figure 3. Place the

board in the enclosure and replace the screws to secure the cell and circuit board.

5. Wait at least four hours for the cell to stabilize before setting the gain.

6. The life expectancy of oxygen cells is about 14 months; warranty is six months. Remove the cell serial number

tag from the plastic bag, and stick it to the inside of the enclosure. In case of cell warranty problems, refer to this
serial number.

N

OTE

: If all channels are used with remote sensor assembly, remove the O2 Cell Circuit Board from the interior of the

ISA-40. Reference Table 2 channel 1.

Remove Shoring
Clip or Spring

Bottom View O2 Sensor

••Center Pin – Negative

••Outer Pin + Positive

O2 Cell Circuit Board
w/cell installed on circuit side

O2 Cell
Circuit Board

See note below