10.0 Troubleshooting

Symptom

Possible Cause

Action

Unit will not turn on.

Dead battery

Plug instrument into charger. Check
charge status.

Sensor fails calibration

Failed sensor

Wrong gas concentration

Old calibration gas

Replace sensor and recalibrate

Use proper calibration gas.

Replace calibration gas.

Calibration due on status line

Sensor hasn’t been calibrated in last 30
days.

Check Cal Status in Pop-Up menu.
Calibrate affected sensor(s).

Cal error on status line

Sensor failed a calibration.

Check Cal Status in Pop-Up menu.
Calibrate affected sensor(s). Replace
sensor if necessary and recalibrate.

Setup/config error on status line

Memory failure

Battery back-up failure.

Return to factory/service center.

Return to factory/service center.

Fault displayed in sensor location

Sensor missing.

Invalid sensor readings.

Properly install sensor.

Check sensor calibration. Replace if
necessary.