Figure 8: various calibration adapter attachments – ENMET TARGET User Manual

Page 17

background image

TARGET

Series

ENMET Corporation

14

Figure 8: Various Calibration Adapter Attachments

Apply the span gas and press the SELECT button to initiate the sample count down timer. Count down is shown on
the lowest line of the display, at the end of the count down the span reading is stored in memory. Remove the span gas
and the calibration shield. Calibration of this sensor is complete.

If the sensor is within acceptable sensitivity parameters, OK is displayed next to the gas momentarily before the
display returns to the Calibration menu.

If there is a problem, Cal error is displayed next to the appropriate channel. See section 10.0 for troubleshooting.

Example:

Display during and after Calibration

Repeat the above procedure for each individual sensor as required. If a calibration error occurs it will also be
displayed as a general condition on the Status Line of the Main Gas Display and can be viewed in the Cal Status
screen of the Pop-Up menu.

Exit the Calibration menu when completed. Remove span gas and calibration shield. Then exit the Standard
Maintenance
menu.

Calibration Adapter, BRH

04834-002

Humidifier Bowl

Regulator

Sensor end cap w/cal shield

Top View of Humidifier Bowl

To Regulator

To Sensor

CAL: O2/CO/H2S/Mth
Sampling MULTI-BLEND
MENU:Abort
Sample in 120 sec.

O2:

OK

CO:

Cal error

H2S:

OK

Mth:

OK

Calibration Adapter, BRH 04834-005

Calibration Adapter, Reactive Gases

Sensor end cap w/cal shield

Cylinder,

Calibration Gas

Regulator

Cylinder, Calibration Gas

Humidifier Bowl

Example of Storage
Case used as Support
for Humidifier

Sensor end cap

w/cal shield

Regulator

Cylinder, Calibration Gas