SPECTRUM Series for Ozone (3E 1 sensor)

The following gases cause a reading:

Interferant

Gas/Vapor

Interferant Concentration

in ppm

Instrument Reading

in ppm

chlorine

1

0.80

fluorine

100ppb

0.07

germane

1

1

nitrogen dioxide

1

0.09

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen sulfide

1

sulfur dioxide

2

phosphine

100 ppb

carbon monoxide

300

carbon dioxide

5000

nitrogen

100%

nitrogen monoxide

10

hydrocarbons

% range

hydrogen

1000

hydrogen cyanide

10

ammonia

10

hydrogen fluoride

3.8

hydrochloric acid

10

SPECTRUM Series for Hydrogen Cyanide (2E 30 sensor)

The following gases cause a reading:

Interferant

Gas/Vapor

Interferant Concentration

in ppm

Instrument Reading

in ppm

hydrogen sulfide

15

40†

chlorine

5

-1

nitrogen dioxide

100

-40†

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen chloride

10*

methane

2000

carbon monoxide

300

carbon dioxide

10 %

freon 12

5000

petrol

300

sulfur dioxide

50*

nitrogen

100 %

nitrogen monoxide

100

cholrinated hydrocarbons

<200

† Long term exposure may destroy the sensor.

*with onboard filter; continuous high level exposure may reduce the filter efficiency