ENMET SPECTRUM ON-LINE User Manual
Page 22

ENMET Corporation
Spectrum ON-LINE
18
SPECTRUM Series for Nitric Oxide (3E 100 sensor)
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
chlorine
5
nitrogen dioxide
100
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
saturated HC, alcohols
abundant
sulfur dioxide
50
SPECTRUM Series for Nitrogen Dioxide (3E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
chlorine
1
3
ozone
1
0.7
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
hydrogen chloride
5
nitrogen oxide
100
hydrogen
1000
carbon monoxide
1000
carbon dioxide
10000
saturated HC, alcohols
abundant
sulfur dioxide
50
ammonia
30
SPECTRUM Series for Hydrazine (2E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
ammonia
200
0.4
methyl hydrazine
1
0.72
1.1 dimethyl hydrazine
1
0.45
chlorine
3
–0.02
hydrogen sulfice
20
0.1
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor
Concentration in ppm
ethanol
1000
Isopropanol
450
hydrogen peroxide
10
hydrogen chloride
7
methanol
1200
hydrogen
2000
carbon monoxide
1000
carbon dioxide
5000
unsaturated HC (ethylene)
1%