ENMET Corporation

Spectrum ON-LINE

18

SPECTRUM Series for Nitric Oxide (3E 100 sensor)

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

chlorine

5

nitrogen dioxide

100

hydrogen

1000

carbon monoxide

1000

carbon dioxide

10000

saturated HC, alcohols

abundant

sulfur dioxide

50

SPECTRUM Series for Nitrogen Dioxide (3E sensor)

The following gases cause a reading:

Interferant

Gas/Vapor

Interferant Concentration

in ppm

Instrument Reading

in ppm

chlorine

1

3

ozone

1

0.7

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

hydrogen chloride

5

nitrogen oxide

100

hydrogen

1000

carbon monoxide

1000

carbon dioxide

10000

saturated HC, alcohols

abundant

sulfur dioxide

50

ammonia

30

SPECTRUM Series for Hydrazine (2E sensor)

The following gases cause a reading:

Interferant

Gas/Vapor

Interferant Concentration

in ppm

Instrument Reading

in ppm

ammonia

200

0.4

methyl hydrazine

1

0.72

1.1 dimethyl hydrazine

1

0.45

chlorine

3

–0.02

hydrogen sulfice

20

0.1

The following gases or vapors have been tested and do not cause a reading:

Interferant Gas/Vapor

Concentration in ppm

ethanol

1000

Isopropanol

450

hydrogen peroxide

10

hydrogen chloride

7

methanol

1200

hydrogen

2000

carbon monoxide

1000

carbon dioxide

5000

unsaturated HC (ethylene)

1%