4.0 Operation

4.1 Activation of the R

ECON

/B

S

ERIES

To activate the detector, press the P

OWER

button

and hold for 3 seconds.

The detector begins a self test sequence, example listed below

1

Display Test

The LCD displays all of the screen elements.

2

Alarm Function Test

The audio alarm beeps, the visual alarm flashes the backlight
activates briefly and the detector emits one vibration

3

Indication of Version

Displays software version

4

Date and Time

Displays the date and time automatically in the following order:
Year, Month & Date, Hour & Minute

5

Indication of Alarm settings

Displays Low Alarm level then High Alarm level

Example: CO

6

Indication of STEL and TWA

For toxic instruments only

Displays STEL then TWA levels

for toxic instruments only

Example: CO

7

Warm Up Count Down

If detector passes the self-test, it enters a short warm up count
down.
3 – 30 seconds, depending on type of gas sensors installed.

Then displays ambient gas reading.

Example: CO in clean air

Proper operation of the R

ECON

/B

S

ERIES

requires that the instrument be zeroed. With the R

ECON

/B

S

ERIES

ON and in clean

air, the display should read 0 or 20.9 for oxygen units. See Section 4.6.

4.2 Deactivation of the R

ECON

/B

S

ERIES

To deactivate the R

ECON

/B

S

ERIES

, press the P

OWER

button

and hold for 3 seconds.

The R

ECON

/B

S

ERIES

will “beep” and display:

Year

Month & Date

Hour & minute

L-alarm

H-alarm

STEL

TWA