5.4 Sensor Replacement

Sensor life in the R

ECON

/B

S

ERIES

is typically: 2 years for Carbon Monoxide & Hydrogen Sulfide and 1year for Oxygen and

Combustible.

The sensor needs to be replaced when the R

ECON

/B

S

ERIES

can no longer be calibrated or if the sensor becomes erratic.

Contact ENMET for sensor replacement information.

Table1: R

ECON

/B

S

ERIES

Factory Set Parameters

Model Gas

Range

Low

Alarm

High

Alarm

TWA

STEL

Cal

Gas

R

ECON

/NH3-B Ammonia

0-100ppm 25ppm

50ppm

25ppm 35ppm

R

ECON

/CO-B Carbon

Monoxide 0-1000ppm 35ppm

100ppm

25ppm 200ppm 500ppm

R

ECON

/H2-B Hydrogen

0-1000ppm

35ppm 250ppm N/A N/A

R

ECON

/HS-B Hydrogen

Sulfide 0-100ppm 10ppm

15ppm

10ppm 15ppm 40ppm

R

ECON

/O2-B Oxygen

0-30%vol 19.5%vol 23.5%vol N/A N/A 20.9%

R

ECON

/SO-B Sulfur

Dioxide

0-100ppm 2ppm

5ppm

2ppm 5ppm

R

ECON

/EX-B Methane

0-100%LEL

10%LEL 50%LEL N/A N/A 50%LEL

R

ECON

/EX-B Propane

0-100%LEL

10%LEL 50%LEL N/A N/A

6.0 Specifications and Troubleshooting Tips

6.1 Specifications

E

LECTRICAL

P

OWER

Working Voltage

3.6V

DC

Lithium battery,1200 mAh Rechargeable

Working time continuous, non-alarm conditions

Combustible gas: Approximately 8 hours

Toxic gas: Approximately 300 hours

Charge time

4 to 6 hours

O

PERATION

Temperature:

–20

C to +50C

Relative Humidity

5 to 95% RH, non-condensing

M

ECHANICAL

Dimensions:

2.4 x 3.9 x 1.3 inches(60 x 100 x 33 mm)

Weight:

5 oz. (140g) including battery

Ingress Protection

IP 65

Approval

Designed to be intrinsically safe

N

OTE

: All specifications stated in this manual may change without notice.

6.2 Troubleshooting Tips

M

ALFUNCTION

P

OSSIBLE

R

EASONS

S

OLUTIONS

Unable to Turn on the detector

low voltage

Charge Battery

Detector Fault

Contact ENMET

Circuit malfunction

Contact ENMET

No reaction to the tested gas

Exhausted Sensor

Replace Sensor

Circuit malfunction

Contact ENMET

Inaccurate indication

Exhausted sensor

Replace the sensor

Overdue Calibration

Calibrate

Wrong time display

Exhausted battery

Charge and reset the time

Electromagnetic disturb

Reset the time

Zero calibration function failed

Too much drift

Calibrate or replace the sensor