0 specifications and troubleshooting tips, 4 sensor replacement, Table1: r – ENMET RECON/B SERIES User Manual
Page 13: Factory set parameters, 1 specifications, 2 troubleshooting tips

R
ECON
/B
S
ERIES
ENMET
11
5.4 Sensor Replacement
Sensor life in the R
ECON
/B
S
ERIES
is typically: 2 years for Carbon Monoxide & Hydrogen Sulfide and 1year for Oxygen and
Combustible.
The sensor needs to be replaced when the R
ECON
/B
S
ERIES
can no longer be calibrated or if the sensor becomes erratic.
Contact ENMET for sensor replacement information.
Table1: R
ECON
/B
S
ERIES
Factory Set Parameters
Model Gas
Range
Low
Alarm
High
Alarm
TWA
STEL
Cal
Gas
R
ECON
/NH3-B Ammonia
0-100ppm 25ppm
50ppm
25ppm 35ppm
R
ECON
/CO-B Carbon
Monoxide 0-1000ppm 35ppm
100ppm
25ppm 200ppm 500ppm
R
ECON
/H2-B Hydrogen
0-1000ppm
35ppm 250ppm N/A N/A
R
ECON
/HS-B Hydrogen
Sulfide 0-100ppm 10ppm
15ppm
10ppm 15ppm 40ppm
R
ECON
/O2-B Oxygen
0-30%vol 19.5%vol 23.5%vol N/A N/A 20.9%
R
ECON
/SO-B Sulfur
Dioxide
0-100ppm 2ppm
5ppm
2ppm 5ppm
R
ECON
/EX-B Methane
0-100%LEL
10%LEL 50%LEL N/A N/A 50%LEL
R
ECON
/EX-B Propane
0-100%LEL
10%LEL 50%LEL N/A N/A
6.0 Specifications and Troubleshooting Tips
6.1 Specifications
E
LECTRICAL
P
OWER
Working Voltage
3.6V
DC
Lithium battery,1200 mAh Rechargeable
Working time continuous, non-alarm conditions
Combustible gas: Approximately 8 hours
Toxic gas: Approximately 300 hours
Charge time
4 to 6 hours
O
PERATION
Temperature:
–20
C to +50C
Relative Humidity
5 to 95% RH, non-condensing
M
ECHANICAL
Dimensions:
2.4 x 3.9 x 1.3 inches(60 x 100 x 33 mm)
Weight:
5 oz. (140g) including battery
Ingress Protection
IP 65
Approval
Designed to be intrinsically safe
N
OTE
: All specifications stated in this manual may change without notice.
6.2 Troubleshooting Tips
M
ALFUNCTION
P
OSSIBLE
R
EASONS
S
OLUTIONS
Unable to Turn on the detector
low voltage
Charge Battery
Detector Fault
Contact ENMET
Circuit malfunction
Contact ENMET
No reaction to the tested gas
Exhausted Sensor
Replace Sensor
Circuit malfunction
Contact ENMET
Inaccurate indication
Exhausted sensor
Replace the sensor
Overdue Calibration
Calibrate
Wrong time display
Exhausted battery
Charge and reset the time
Electromagnetic disturb
Reset the time
Zero calibration function failed
Too much drift