5.1.3 Calibration Point Adjustment

In this mode, display is the last concentration of gas used to calibrate the

instrument. Press UP

or Down

button to modify the display.

N

OTE

: Not used with Recon Oxygen units

N

OTE

: 20.9% Oxygen is the only calibration level to be used with Recon Oxygen units

Open the valve on the regulator

Within 30 seconds, if the R

ECON

/B

S

ERIES

detects the calibration gas, it will

start up the calibration date analysis procedure by itself. Continue gas flow.

The Display

will indicate gas level

as gas is detected

In 2 to 3 minutes the R

ECON

/B

S

ERIES

will automatically adjust the conversion

arithmetics to complete the gas response process. Then the R

ECON

/B

S

ERIES

saves the best arithmetics to complete calibration.

W

ARNING

:

The R

ECON

/B

S

ERIES

must self calibrate, please avoid

touching any buttons, otherwise the accuracy will be effected badly.

After that it enters into the L-alarm set up. Press P

OWER

button twice to

advance and exit.

If the R

ECON

/B

S

ERIES

can not reach half the calibration figure within 30 seconds, or the gas concentration is outside the

maximum drift range of the sensor, the E icon will light and 0000 will be displayed. This means the gas is not proper for
calibration or the sensor is depleted.

Verify calibration gas is correct. If correct retry calibration sequence. If failure occurs sensor may need to be replaced.

Remove calibration cover from instrument.

To exit calibration menu.

Press S

ET

button 3 times and then the P

OWER

button 4 times.

Press the S

ET

button then

Press P

OWER

button