Guidance and manufacturer’s declaration – electromagnetic immunity
The model AID-173SHR is intended for use in the electromagnetic environment specified below. The
customer or the user of the model AID-173SHR should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
– guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the model
AID-173SHR, including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1,2 P
d = 1,2 P 80 MHz to 800
MHz
d = 2,3 P 800 MHz to 2,5
GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation