Guidance and manufacturer’s declaration – electromagnetic immunity

The model AID-173SHR is intended for use in the electromagnetic environment specified below. The customer or
the user of the model AID-173SHR should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic

environment

guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

6 kV contact

8 kV air

6 kV contact

8 kV air

Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.

Electrical fast
transient/burst

IEC 61000-4-4

2 kV for power

supply lines

1 kV for input/output

lines

2 kV for power

supply lines

1 kV for input/output

lines

Mains power quality should
be that of a typical
commercial or hospital
environment.

Surge
IEC 61000-4-5

1 kV line(s) to line(s)

2 kV line(s) to earth

1 kV line(s) to line(s)

2 kV line(s) to earth

Mains power quality should
be that of a typical
commercial or hospital
environment.

interruptions and
voltage variations
on power supply
input lines

IEC 61000-4-11

<5 % UT
(>95 % dip in UT)
for 0,5 cycle

40 % UT
(60 % dip in UT)
for 5 cycles

70 % UT
(30 % dip in UT)
for 25 cycles

<5 % UT
(>95 % dip in UT)
for 5 sec

<5 % UT
(>95 % dip in UT)
for 0,5 cycle

40 % UT
(60 % dip in UT)
for 5 cycles

70 % UT
(30 % dip in UT)
for 25 cycles

<5 % UT
(>95 % dip in UT)
for 5 sec

Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of the
model AID-173SHR requires
continued operation during
power
mains interruptions, it is
recommended that the model
AID-173SHR be powered
from an uninterruptible power
supply or a battery.

Power frequency
(50/60 Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.