Ambient conditions – VACUUBRAND RC 6 User Manual
Page 11
page 11 of 59
(e.g., housing, walls, etc.), or else install an external au-
tomatic ventilation system. Check fan regularly for dust/
dirt. Clean fan guard if necessary to avoid a reduction of
ventilation.
The On/Off switch is located at the side of the terminal
box.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s connections (to avoid overpressure at the outlet,
and reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all
applicable and relevant safety require-
ments (regulations and guidelines). Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
➨
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+
Do not use this product in an area where it can fall or
be pulled into water or other liquids.
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, installation in higher altitudes, conductive
pollution or external condensation on the pump.
• Do not operate this product near flames.