If pumping condensable vapors (water vapor,
solvents, etc.), let the pump run with
gas bal-
last to help purge any condensation in the
pump.
gas ballast
(open)
Prevent internal condensation, transfer of liq-
uids or dust. The diaphragms and valves will
be damaged, if liquids are pumped in signifi-
cant amounts.
Check the pump regularly for external soiling
and deposits. Clean the pump if necessary
to avoid an increase of the pump’s operating
temperature.
NOTICE
IN
✘
condensate / dust
In case of overload, the motor is shut down by
a
self-hold thermal cutout in the winding.
Note: Only manual reset is possible. Switch
off the pump and disconnect from the power
source. Identify and eliminate the cause of fail-
ure. Allow the pump to cool down sufficiently
before restart.
°C / °F
I ➨ O
!
3.
2.
1.
IN
✘
Avoid overheating (e.g., due to hot process
gases). Maximum permissible temperature
range: see “Gas inlet temperatures”, pg. 27.
Make sure the fans’ air supply is adequate.
Do not place the pump on soft surfaces (e.g.,
rubber foam) during operation. This may re-
duce the air circulation in the pump housing
and affect the fans’ cooling effect. Check fans
regularly for dust/dirt. Clean fan guard grills if
necessary to avoid a reduction of ventilation.