Technical data, General technical data valid for all pumps, Protection, “technical data”, pg. 24) – VACUUBRAND MV 10C NT VARIO o.C. User Manual
Page 24
![background image](/manuals/715219/24/background.png)
page 24 of 97
Technical data
General technical data valid for all pumps
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum pressure difference between
inlet and outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure (abso-
lute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric moisture
during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
ft
(m)
6500
(2000)
Rated motor power (pump motor)
hp
(kW)
0.71
(0.53)
No-load speed
rpm
30 - 2400
Maximum permissible range
of supply voltage (±10%)
Attention: Observe specifications of rating plate!
100-120 V~ 50/60 Hz,
200-230 V~ 50/60 Hz
Maximum rated current at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
8.0
3.5
Device fuse
2 slow blow fuses 8A
Motor protection
thermal cutout, manual reset
Degree of protection IEC 529
IP 40