Error codes – Triner Scale TS 700-SS User Manual

Page 19

background image

17

Error Codes

ERROR

REASON

SOLUTION

UUUUUU

1. Overload
2. wrong connection
with load cell
3. load cell has quality
problem.

1. reduce the weight
2. check load cell connection
3. inspection load cell. Check
the input and output

nnnnnnn

1. calibration is no
good
2. wrong connection
3. load cell has quality
problem

1. check scale is resisted or

not, foot is kept level or not.

2. check load cell connection.
3. checking load cell

check

input and output resistance to
judge it is good or not.

ERR1


During calibration, not
input the weights or
the weight is overload

Input the correct weights

ERR2


During calibration , the
weights is below than
Min. required weights

The

calibration

weights

Minimum is 10% of Max. cap.
Recommend 60%-80% of Max.
Cap.

ERR3


During calibration, the
input

signal

is

negative

1. check the connection is

correct

2. check load cell is no

problem

3. recalibration if still wrong

change the PCB

ERR4

During calibration, the
signal is unstable

After the platform is stable,
start calibration

ERR5

Change PCB