Certificate of conformance – ThermoWorks CF-IR CLOSE FOCUS INFRARED THERMOMETER User Manual

Page 2

background image

CAUTION

1. Do not look directly into the laser beam. Permanent eye damage may result.
2. Never point the device towards anyone’s eyes.
3. Keep out of reach of all children.

STORAGE & CLEANING

The sensor lens should be kept clean at all times and stored at room temperature. Use only a soft cloth or cotton swab

with mater or medical alcohol when cleaning the lens. Allow the lens to fully dry before using. Clean the thermometer

with a damp cloth. The thermometer should be stored between -4 to 149°F / -20 to 65°C.
Readings may be affected if the thermometer is operated within a radio frequency electromagnetic field strength of

approximately 3 volts per meter, but its performance will not be permanently affected. Under an electromagnetic field of

3V/m from 300 to 350 mHz the maximum error is ±6°C.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as follows:

Er2 is displayed when the thermometer is exposed to rapid changes in the ambient temperature.

Er3 is displayed when the ambient temperature exceeds 0°C (32°F) or +50°C (122°F). The thermometer

should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.

Er is displayed for all other error messages. Clear the error by resetting the thermometer. To reset, turn the

thermometer off, remove the batteries and wait for a minimum of one minute. Reinsert the batteries and turn

it on. If the error message remains please contact ThermoWorks for further assistance.

Hi or Lo is displayed when the temperature being measured exceeds the measurement range of the

thermometer.

BATTERIES

The thermometer incorporates visual low battery indication as follows:

‘Battery OK’

‘Battery Low’: battery

needs to be replaced.

‘Battery Exhausted’:

measurements may be

affected.

Utah, USA

Ph: 801-756-7705

Fax: 801-756-8948

www.thermoworks.com

CERTIFICATE OF CONFORMANCE

The manufacturer of this instrument has implemented a quality assurance

system under ISO 9001:2000 certified quality system and fully follow

ISO GUM (Guide to the Expression of Uncertainty in Measurement)

to evaluate the uncertainty of temperature and resistance standards,

guarantee performance as below:

Calibration

Temperature

Max

Error

Target

Accuracy

Target

Stability

-30°C

±2.0°C

±0.3°C

0.3°C

0°C

±2.0°C

±0.4°C

0.1°C

100°C

±2.0°C

±0.4°C

0.1°C

200°C

±4.0°C

±0.5°C

0.1°C

Note: Assume the operation ambient temperature under 23 ±3°C

Furthermore, we certify that this infrared thermometer has been

inspected and found to comply with published specifications. This

device has been calibrated by temperature and/or resistance standards

that are traceable to NIST (U.S. National Institute of Standards and

Technology) or NML (National Measurement Laboratory) of Taiwan,

and the calibration procedure corresponds with generally accepted

regulations and standards.

ThermoWorks, Inc.

1762 W. 20 S., #100

Lindon, UT 84042

www.thermoworks.com