INSTRUMENT OPERATION

The Instrument is switched on and off by using the ‘ON/OFF’ button located on

the front of the instrument.

Apply the tip of the probe to the substance, medium or surface to be measured.

The instrument reading may take seconds to stabilise depending on the nature

of the measurement and the sensitivity of the probe.

SENSORS

Therma Plus instruments should only be used with a ‘Type K’ thermocouple

probe made to BS EN 60584:1996 that is fitted with a suitable miniature ther-

mocouple plug. Go to www.thermoworks.com for suitable probes that measure

liquids, air, solids, and surface temperatures. Custom probes are also available.

HOLD

Pressing the ‘HOLD’ button freezes the display for ease of reading and also

illuminates the hold segment on the display. Pressing ‘HOLD’ again returns the

instrument back to normal operation.

UNITS

The display units can be changed to read in °F or °C. To change the current

display units first turn the unit off. While pressing and holding down the

HOLD button turn the meter on. Keep the HOLD button pressed (about 7

seconds) until the units change on the display. Release

HOLD button.

RESOLUTION

The instrument automatically changes range from 0.1º measurements to 1º

measurements after 299.9ºC or 599.9ºF respectively.

AUTO-OFF

The instrument will turn off 20 minutes after the last button press.

UNDER-RANGE

If you are attempting to measure below the instruments specified temperature

range, LO will be displayed.

OVER-RANGE

If you are attempting to measure above the instruments specified temperature

range, HI will be displayed.

OPEN-CIRCUIT

If no probe is fitted (Therma Plus) or there is a fault with the probe then the

display reading will rise until eventually displaying Err.

BATTERY REPLACEMENT

LO BAT indicates that the batteries will need replacing soon. The instrument con-

tinues to measure accurately while in that ‘Lo Bat’ condition but we recommend

that the battery be changed at the earliest convenience. After some further

usage the instrument will detect that there is insufficient battery voltage left

to measure accurately. At this point the instrument will display ‘Flat’ and then

‘Bat’ and then shut down. The batteries will need to be replaced at this time to

continue instrument use.

To change the battery, remove the instrument bottom cover using a crosshead

(Phillips) screwdriver. Once opened push the battery to one side to release

and remove it and then fit the new battery into the battery clips taking into

account the polarity. The battery should be a PP3, MN1604 or a direct 9-volt

equivalent.

When refitting the bottom cover ensure that the seals are not damaged and are

correctly positioned. Tighten the battery cover screws firmly to ensure that the

seals are compressed and will provide a water-resistant seal.

EMC/RFI

Readings may be affected if the unit is operated within a radio frequency

electromagnetic field strength of approximately 3 volts per metre, but the

performance of the instrument will not be permanently affected.

ENCLOSURE

The enclosure is manufactured from extruded aluminum and is rated to

IP66/67 water resistant rating.

REPLACING PROBES (CaterTemp)

The hard-wired probes in the CaterTemp can be easily replaced if damaged. We

recommend returning the unit to ThermoWorks for this probe replacement. At

that time we will recalibrate the CaterTemp with the new probe attached so your

“system accuracy” is within the original specification. Probes can be replaced

in the field by the user if you remove the top end cap and unplug the probe

from the circuit board. However, your unit may not meet the original “system

accuracy” specification unless a “system calibration” is performed. Contact

ThermoWorks for further questions.

PRODUCT WARRANTY & GUARANTEE

This instrument carries a two-year warranty and guarantee against defects in

either components or workmanship. In cases where the instrument has a fixed

probe then the length of the warranty for the probe is 3 months. During this

period, products that prove to be defective will, at the discretion of Thermo-

Works, be either repaired or replaced without charge. The product warranty

and guarantee

does not cover damage caused by fair wear and tear,

abnormal storage conditions, incorrect use, accidental misuse,

abuse, neglect, misapplication or modification, or use with non-Thermo-

Works hardware/software. No warranty of fitness for a particular purpose is

offered and the user assumes the entire risk of using the product. Any liability

by ThermoWorks is limited only to the replacement of defective materials or

workmanship, and ThermoWorks accepts no responsibility for consequential

loss. In line with our policy of continuous development, we reserve the right to

amend our product specification without prior notice.