Power – Shellab SSI10R-2 User Manual
Page 19

SHEL LAB SI9/SI9R Shaking Incubator Operational Manual
19
Copyright © 2007 Sheldon Manufacturing. All rights reserved.
All specifications and features for this product are subject to change without notice.
What is the problem?
Possible Causes
To solve the problem...
Cannot adjust set points or
calibration.
This is a controller failure.
1. Turn entire unit off and then
on to reset the unit. This
may temporarily solve the
problem, but controller may
be faulty.
If this does not solve the
problem...
2. Call customer service. See
Power
What is the problem?
Possible Causes
To solve the problem...
The unit will not turn on.
Power
cord
not
firmly
plugged into the outlet.
The unit or wall fuse/circuit
breaker has blown.
The outlet is defective.
The unit is plugged into a
circuit that already has too
many electrical loads.
1. Be sure the voltage and
frequency specifications of
the outlet are within the
range stated on the power
rating overlay at the rear of
the unit.
If this does not solve the
problem...
2. Check the power cord at the
electrical outlet for proper fit.
3. Make sure the unit is
plugged in firmly.
If this does not solve the
problem...
4. Replace fuse/circuit breaker
in the unit or wall as
necessary.
If this does not solve the
problem...
5. Make sure the outlet is in
proper working condition.
6. Replace if defective.
If this does not solve the
problem...
7. Check to see what other
loads are on the same circuit
as the unit. We recommend
that you plug your incubator
into a circuit separate from
other equipment.
8. Call customer service. See
The unit fuse/circuit breaker blows
often.
Wrong fuse installed.
Wire is shorting out.
1. Check fuse for right
amperage.
2. Call customer service. See