Solving Problems

Should the proper function of your unit come into
question, use this section to help you determine what
the problem is and how to fix it.

Check if your question is similar to those listed below.
Then follow the guidelines found in that section:

The temperature control inside the unit does not
appear to be working correctly. What’s wrong?

The refrigeration of my SSI5R (SI6R) does not
appear to be working correctly. What’s wrong?

WARNING

Replacing fuses with wrong type and value can
result in serious damage to the equipment and
property. ONLY Replace fuses with the same type
and amperage indicated on the replacement fuse
labels.

There are no user serviceable components inside
the unit. Potentially lethal voltages exist. Do not
attempt to service your unit beyond the procedures
described here. See “Getting Your Unit Serviced on
page-21

Temperature

The temperature inside the unit is difficult to control. What’s wrong?

What is the problem?

Possible Causes

To solve the problem...

The temperature indicated by the
Main Temperature Control is
higher than my reference
thermometer.

Controller is calibrated too

high.

1. Calibrate the Main

Temperature Controller. See
page 11.

2. Call customer service. See

page 21.

Display reads “HI” or “400+”.

Probe is unplugged

- Or -

Wire to the probe is broken.

- Or -

Probe is plugged in

backwards.

1. Be sure the temperature

probe is properly plugged in.

If this doesn’t solve the
problem...

2. Call customer service. See

page 21

Chamber temperature spikes over
the set point.

Unit is not calibrated
properly.

1. Calibrate the Main

Temperature Controller. See
page 11.

The temperature indicated by the
Main Temperature Control is
lower than my reference
thermometer.

The temperature inside the
unit has not yet stabilized
after the door has been
opened.

- Or -

The temperature inside the
unit has not yet stabilized
after the unit has been
turned off or a power failure.

- Or -

Controller is calibrated too
low.

- Or -

1. Wait for the temperature

indicated by the Main
Temperature Controller to
stabilize. If you have just
turned the unit on, wait 24
hours for the incubator to
stabilize at a warmer
temperature. A fluctuation of
no more than + 0.1

C is

normal.

If this is not the problem...

2. Recalibrate the Main

Temperature Controller. See
page 11.