Refrigeration – Shellab SRI21D-2 User Manual
Page 16

16
1/ ±0.1 may be normal 
2/ is fan working? 
– fan must operate for uniformity
3/ is ambient room temperature radically changing
– either door
opening or room airflow from heaters or air conditioning ?
–
stabilize ambient conditions. 
4/ sensor miss-located, damaged or wires may be damaged - 
check mounts for control and Thermostat sensors, then trace wires 
or tubing between sensors and controls. 
5/ calibration sensitivity 
– call Customer Service
6/ Overtemperature set too low
– be sure that Thermostat is more
than 5 degrees over desired set point; check if OVERTEMP pilot is 
on continuously; turn controller knob completely clockwise to see if 
problem solved then follow instructions in owner’s manual for 
correct setting 
– see section 5.8
7/ electrical noise
– remove nearby sources of RFI including
motors, arcing relays or radio transmitters. 
8/ bad connection on temperature sensor or faulty sensor 
– check
connectors for continuity and mechanical soundness while 
watching display for erratic behavior; check sensor and wiring for 
mechanical damage. 
9/ bad connections or faulty relay 
– check connectors for
mechanical soundness and look for corrosion around terminals or 
signs of arcing or other visible deterioration. 
Display and reference thermometer 
don’t match 
1/ calibration error
– see section 5.9
2/ controller failure
– evaluate if pilot light is operating correctly
3/ allow at least two hours to stabilize. 
4/ see if reference thermometer is certified. 
Calibrated at one temperature, but 
not at another 
Refer to Section 5.9
REFRIGERATION
Temperature can't get up to set point
1/ assure that power is going to heating coils. 
2/ confirm that evaporator is calling for heat (check front panel 
light).
3/ if light not coming on, check control set point and Thermostat
set point .
4/ confirm that fan is operating and airflow is not blocked. 
5/ reset by turning unit off and on. 
Unit won't cool