Lbl g, None, Set 2 – Shellab SRI21D-2 User Manual

Page 11: Setting and calibrating night cycle, Value, I.ca, Oper, Usr.s

background image

11

BEFORE CALIBRATION IS MADE MAKE SURE THAT INDICATOR 2, 3, 5, AND RAMP SYMBOL ARE ON INDICATING DAY

CYCLE ACTIVE. FIGURE OUT # VALUE OF OFFSET (ACTUAL TEMPERATURE MINUS PROCESS DISPLAY TEMPERATURE).

BELOW SHOWS WHERE AND HOW TO ENTER THE OFFSET # VALUE.

20.0

20.0

PUSH

AND

HOLD

FOR

5

SECONDS

PUSH

THREE

TIMES

i.CA

# VALUE

USE

UP AND

DOWN

BUTTONS

TO

ENTER

CALIBRATION

# VALUE

i.CA

# VALUE

PUSH

TWICE

20.0

20.0

TO RETURN

TO NORMAL

DISPLAY

A1

oPEr

OR UNTIL

SCREEN

APPEARS

AFTER THE OFFSET HAS BEEN ENTERED ALLOW UNIT TO STABALIZE. RECHECK TEMPERATURE AGAIN
AND IF CALIBRATION IS WITHIN EXCEPTABLE LIMITS THE OFFSET # VALUE NEEDS TO BE STORED INTO

MEMERY OR IT WILL REVERT BACK TO THE LAST # VALUE ON THE NEXT SWITCHING CYCLE.

BELOW SHOWS HOW TO STORE OFFSET # VALUE FOR THE DAY CYCLE

A1

SET

20.0

20.0

PUSH

AND

HOLD

FOR

10

SECONDS

PUSH

THREE

TIMES

LbL

g

SET

PUSH

FOUTEEN

TIMES

OR UNTIL

SCREEN

APPEARS

nonE

Usr.S

Usr.S

SET 2

PUSH

ONCE

PUSH

TWICE

20.0

20.0

TO RETURN

TO NORMAL

DISPLAY

SELECTING SET 2 UNDER Usr.S SAVES THE OFFSET CALIBRATION FOR THE DAY CYCLE.

SETTING AND CALIBRATING NIGHT CYCLE

Use the up/down arrows to choose desired Setpoint. After Setpoint is entered, turn ON both Event Outputs.
To simulate Day Cycle, turn Event Outputs OFF to simulate Night Cycle. To turn OFF the Event Outputs,
follow directions below.

Outputs 2,3, and 5 should be OFF and 6 should be illuminated in the bottom of display. The lights inside
incubator should be off. Allow unit to stabilize for several hours before calibration.