beautypg.com

Mechanical, Other – Shellab SGO3-2 User Manual

Page 13

background image

13

1/ ±0.1 may be normal.
2/ For G suffix models: may vary ±2.0 degrees.
3/ For F suffix models: is fan working?

–verify movement of air in

chamber.
4/ Is ambient room temperature radically changing

– either door

opening or room airflow from heaters or air conditioning ?

– stabilize

ambient conditions.
5/ This may happen if exhaust stack is 100% open or if power exhaust
is cycling

– adjust stack to at least ¼ closed.

6/ Sensor miss-located, damaged or wires may be damaged - check
mounts for control and OTP sensors, then trace wires or tubing
between sensors and controls.
7/ Calibration sensitivity

– call Customer Service.

8/ High limit set too low

– be sure that Safety Thermostat is set more

than 5 degrees over Main Controller set point; check if OTP pilot is on
continuously; turn controller knob completely clockwise to see if
problem solved then follow instructions in section 6.5 for correct setting.
9/ Electrical noise

– remove nearby sources of RFI including motors,

arcing relays or radio transmitters
10/ Bad connection on temperature sensor or faulty sensor

– check

connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage.
11/ Bad connections or faulty solid state relay

– check connectors for

mechanical soundness and look for corrosion around terminals or signs
of arcing or other visible deterioration.

Will not maintain set point

1/ Assure that set point is at least 5 degrees over ambient.
2/ See if ambient is fluctuating; check for adjacent open doors or
HVAC duct openings

– stabilize ambient conditions.

Display and reference thermometer don’t
match

1/ Calibration error

– see section 6.4.

2/ Temperature sensor failure

– call Customer Service.

3/ Controller failure

– call Customer Service.

4/ Allow at least two hours to stabilize.
5/ Verify that reference thermometer is certified.

Can't adjust set points or calibration

1/ Turn entire unit off and on to reset.
2/ If repeatedly happens, call Customer Service.

Calibrated at one temperature, but not at
another

This can be a normal condition when operating temperature varies
widely. For maximum accuracy, calibration should be done at or as
close to the set point temperature.

MECHANICAL

Motor doesn't move; F suffix models

1/ If shaft spins freely: check connections to motor and check voltage to
motor.
2/ If shaft rubs or is frozen, relieve binding and retest.

Motor makes noise; F suffix models

1/ If noise is from the motor, tap the top of motor shaft with ball peen
hammer.
2/ If the sound gets worse, tap the other end of the shaft

– avoiding

touching the blower wheel.
3/ If there is no change, call Customer Service.

Door not sealing

1/ Adjust hinge blocks or twist the door.
2/ Confirm that unit has not been damaged and body is not square.
3/ Check physical condition of gasket for tears or punctures.

OTHER

This manual is related to the following products: