Shellab SCO2W-2 User Manual
Page 19

19
1/ turn set point up and down to see if solenoid valve works by feeling 
and listening to valve 
2/ recalibrate with Fyrite, see section 5.11 and section 7.0 
3/ fan not operating correctly 
 a- fan motor stopped 
 b- fan blade fell off 
 c- wrong fan blade installed or mounted backwards 
4/ wrong restrictor installed or missing altogether 
5/ tank pressure too high, see section 5.9.A 
6/ CO2 sensor partially plugged with dirt or condensation 
7/ regulator set wrong, see section 5.9 
8/ incubator too heavily loaded 
9/ incubator being operated without shelving 
10/ CO2 attached to sample port instead of “CO2 IN” fitting 
12/ if display and Fyrite does not match, see Troubleshooting Co2-5 
Overshoots set point and continues to rise 
- display and Fyrite match 
1/ debris in solenoid causing it to leak continuously 
2/ solenoid failed while open 
3/ unit miss-wired 
4/ controller output failed or shorted 
5/ CO2 sensor or interface failure 
6/ CO2 sensor plugged by debris or condensation 
Rises very slowly
1/ restrictor partially plugged 
2/ filter overly dirty or partially plugged 
3/ CO2 tank regulator set too low, see section 5.9 
4/ hose kinked or leaking 
5/ poor door seal 
6/ CO2 tank contains mixed gas, not 100% medical grade CO2 
Never rises
1/ CO2 tank empty 
2/ solenoid failed while closed 
3/ CO2 controller output failed while open 
4/ restrictor plugged 
5/ CO2 hose blockage 
6/ CO2 filter plugged 
7/ set point is at 0.0 and has not been reset, see section 5.3 and 5.4 
Display and Fyrite reading do not match
1/ calibration error - clear chamber for 12 hrs and confirm at "0" 
2/ turn set point up and down to see if solenoid valve works by feeling 
and listening to valve 
3/ assure that IR tube block is completely tight, 24 series 
4/ if display is varying when doing Fyrite test, have a leak on IR base or 
mounting plate, or CO2 is injecting at time of test. 
5/ take voltage reading across sensor leads 0=0V and 20=1V 
6/ assure that sensor is properly mounted 
7/ DISPLAY LOWER THAN FYRITE: air leak around the CO2 sensor or 
the fan motor shaft 
8/ DISPLAY HIGHER THAN FYRITE: leak in tube from sample port to 
chamber, verify with liquid soap solution. 
9/ DISPLAY HIGHER THAN FYRITE: Fyrite fluid worn out, replace fluid 
with Fyrite refill. 
10/ DISPLAY HIGHER THAN FYRITE: poor Fyrite sampling technique, 
see section 6.0 
11/ DISPLAY HIGHER THAN FYRITE: defective Fyrite pump or tubing 
12/ CO2 sensor, interface or controller failure 
Is unstable
– display or actual reading
varies around set point