LABEC Z-ZHWY-1112B User Manual
Page 8
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
10. Non-volatile Memory Function
Non-volatile memory retains parameters during a power interrupt and restarts the
incubator shaker automatically. Non-volatile Memory Function can be set by the
following steps:
1. First press the
Set/Confirmation
key, then press the
Increase
key to enter the
password 5.
2. Press the
Timer
key, the unit enters into non-volatile memory setting mode.
The LED indicator displays ON/OFF options.
3. Press the
Increase/Decrease
key to choose between ON and OFF. When ON is
selected, the non-volatile memory function is started. When Off is selected,
the memory function is canceled.
11. Technical Specifications
Item
Incubator
Shaker
Incubator
Shaker
Incubator
Shaker
Incubator
Shaker
Model
ZHWY-2102
ZHWY-2102C
ZHWY-2112B
ZHWY-2112F
Performance
Heated and Refrigerated
Heated and Refrigerated
Temperature
Range
Ambient: -20°C to 60°C
Ambient: -20°C to 60°C
Temperature
Set Division
±
0.1°C
±
0.1°C
±
0.1°C
±
0.1°C
Temperature
Uniformity
±
1°C
±
1°C
±
1°C
±
1°C
Shaking Speed
60-300 rpm
(±1rpm)
60-300 rpm
(±1rpm)
60-300 rpm
(±1rpm)
60-300 rpm
(±1rpm)
Timer
0-500 hours
0-500 hours
0-500 hours
0-500 hours
Capacity
50mlx90 or
100mlx90 or
250mlx56 or
500mlx48 or
750mlx36 or
1000mlx30
50mlx48 or
100mlx48 or
250mlx30 or
500mlx24 or
750mlx18
50mlx154 or
100mlx154 or
250mlx90 or
500mlx70 or
750mlx56 or
1000mlx36
50mlx154 or
100mlx154 or
250mlx90 or
500mlx70 or
750mlx56 or
1000mlx36
Platform Size
734x458 mm
500x350 mm
970x560 mm
940x580 mm
Overall Size
700x590x1230
mm
950x710x1250
mm
1200x820x1600
mm
1200x820x1600
mm
Electrical
Requirement
240V 50Hz
240V 50Hz
240V 50Hz
240V 50Hz
Weight
235Kg
150Kg
456Kg
486Kg
Warranty
1 year