LABEC Z-ZHWY-1112B User Manual
Page 6
![background image](/manuals/697882/6/background.png)
Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01
Refrigeration
Press the
Set/Confirmation
key, then press the
Increase
key to enter the password 9.
Press the
Refrigeration
key, the unit enters into refrigeration setting mode.
Refrigeration parameter can be set via the
Increase/Decrease
key. To confirm the new
setting, press the
Set/Confirmation
key.
•
If the unit is to be used in the temperature range of ambient-5°C to ambient+5°C, the
refrigeration parameter can be set at the value 0.2.
•
If the unit is to be used in the temperature range of ambient-20°C to ambient-5°C, the
refrigeration parameter can be set at the value 0.5.
•
If the unit is to be used in the temperature range of ambient+5°C to 60°C, the
refrigeration parameter can be set at the value 2. The refrigeration is cancelled.
6. Overtemperature Alarm
In case the gap between measured temperature and setting temperature excels ± 3°C,
the unit sends out audible and visible alarm automatically, and the yellow alarm
indicator flashes. The platform keeps on shaking while the heating or refrigeration
stops. Press the
Temperature
key will cancel the alarm for 30 minutes. Reset the
temperature value, the unit exits the alarm mode.
7. Speed Control
To set the desired shaking speed, first press the
Set/Confirmation
key, then press the
Increase
key to enter the password 3. Press the
Speed
key, the unit enters into speed
setting mode. The desired speed value can be set via the
Increase/Decrease
key. To
confirm the new setting, press the
Set/Confirmation
key. Setting of a new value
overwrites the old value in the memory.
8. Timer
When timer function is needed, first press the
Set/Confirmation
key, then press the
Increase
key to enter the password 3. Press the
Timer
key, the unit enters into timer
setting mode, the desired timer value can be set via the
Increase/Decrease
key. To
confirm the new setting, press the
Set/Confirmation
key.
By the time the remaining time reaches 0:00, the platform stops shaking, the unit sends out audible
and visible alarm for 2 minutes. Press the
Timer
key, the alarm will be muted.