Appendix, Form1 cycle mode chart, Form2 work mode chart – LABEC AA-12S User Manual
Page 27: Sterilize mode, as following form(8l/12l), Operation manual

SEA Series Steam Sterilizers
Operation manual
2 3
Appendix.
Form1 Cycle mode Chart
Item
Digital display
Mode
Description
1
Ld
Pre-heating
Preparation and conditioning of the
autoclave
2
HE
Heating
Chamber is being heated and pressurized
3
Cycle time
Sterilizing
Counting down of sterilizing
4
Dr_
Venting
Chamber starts to vent and dry
5
Drying time
Drying
Instruments are being dried.
6
Ed_
Complete
Cycle is complete and chamber door can
be opened
7
Faults & Error codes
Error codes
Displays error codes, resets and alarm
8
LE_
Balance state
When change the state, the pressure of inside
and outside the Autoclave should be the same.
Form2 Work mode chart
Sterilize mode, as following form(8L/12L)
o
134 C
package
o
134 C
non package
o
121 C
package
o
121 C
non package
134B+
BD test
Vacuum
test cycle
CLEAN test
Sterilizing
temperature
134
134
121
121
134
134
Sterilizing pressure
2.10
2.10
1.10
1.10
2.10
2.10
Classes
S
S
S
S
S
B
Pre-heating min
0-7
0-7
0-7
0-7
0-7
0-7
Vaccumize time min
4
4
4
4
4
10
4
Clean 20
Admission time
min
5
5
5
5
5
5
Drying
Sterilizing time
min
4
6
16
20
18
3.5
Keep
pressure 10
Drying time
min
10
10
10
10
10
1
10
Quick drying time
1/5
1/5
1/5
1/5
1/5
Total time
min
30(21/25)
38(29/33)
42(33/37)
52(43/47)
50(41/45)
26.5
14
30