KERN & SOHN MPB 300K100P User Manual
Page 14

MPB / MXS -BA-e-1113
14
5.6.1 Procedure when adjusting (only dealers authorised by KERN)
Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization.
Operation
Display
Turn on balance by pressing the [ON/OFF] key
Keep [
0
] button depressed for approx. 3 sec until
"SETUP" is displayed followed by "9rAd"
Press the [ TARE ] button as often as required until
"CAL ib" is displayed
Press the [HOLD] button.
Press the [TARE] key. ( Zero count mode )
Press the [
0
] button repeatedly until "CAL 0" is
displayed.
Press the [HOLD] button.
Press the [TARE] key.
Enter the required size of the adjustment weight (See
Section 1, "Technical Data"): Press the [HOLD] button
to select the item to be changed and press the [TARE]
button to select the numerical value.
Confirm by pressing the [
0
] button.