9 error messages – KERN & SOHN HFT 15T5 User Manual
Page 37

HFT-BA-e-1310
37
9 Error messages
Error
message
Description
Possible causes
--ol-
Maximum load exceeded
Reduce load
Check whether the balance has
been damaged
Err 4
Zeroing range exceeded due
to switching-on balance or
pressing
(normally 4%
max)
Load attached
Overload during setting to zero
Err 5
Keyboard error
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Damaged weighing cell
Damaged electronics
Err 10
No signal
Distance emitter / receiver too
long
Emitter (connecting box) not
switched on
Lo bA
Capacity of rechargeable
battery exhausted
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error