beautypg.com

KERN & SOHN BBB 6T-3M User Manual

Page 15

background image

KFP V20_600-6000-IA-e-1313

15

9 Service documentation

This chapter is only intended for a balance specialist!

The weighing bridges are carried out in DMS sensor technology, at
every corner a DMS weighing cell is installed.

The analogue-digital transformation occurs in the display unit. Also all
the balance and country-specific data are stored there.


9.1 Overview, setting regulation, tolerances

Testing and setting regulations:

Capacity

600 kg

1500 kg

3000 kg

6000 kg

Readability

200 g

500 g

1000 g

1000 g

Min

4 kg

10 kg

20 kg

40 kg

Max

600 kg

1500 kg

3000 kg

6000 kg

1/3 corner load

200 kg

500 kg

1000 kg

2000kg

Tolerance

200 g

500 g

1000 g

2000g

This manual is related to the following products: