3 preload, deadload and overload settings – KERN & SOHN KFP V20 User Manual
Page 9

6.3 Preload, Deadload and Overload settings
Kern model
max. Preload* (kg)
* = additional initial
load
Deadload** (kg)
**= initial load placed
earlier
Center Overload
Protection circa (kg)
Corner Overload
Protection circa (kg)
Loadcell
Capacity (kg)
KFP 6V20 M
2.28
1.72
8.5
5
10
KFP 6V20 LM
1.86
2.14
8.5
5
10
KFP 15V20 M
2.86
2.14
23
12
30
KFP 30V20 M
10.52
4.48
46
30
50
KFP 60V20 M
35.52
4.48
85
50
100
KFP 60V20 LM
30.98
9.02
85
50
100
KFP 150V20 M
90.98
9.02
200
130
200
KFP 150V20 LM
136.14
13.86
270
130
300
KFP 300V20 M
186.14
13.86
550
230
500
KFP600V20AM
340
22.8
900
450
750
Platform type
Platform
dimension
(mm)
Loadcel
l
TC
Class
Max
E
max
E
min
Y
V
min
n
Dead-
load
T
min
T
max
Z
Cable- P
Lc
Typ
Nr.
Preload
-1
-4
-2
-3
(kg)
-5
-6
oder length
(kg)
(kg)
(g)
(g)
DR
(m)
KFP 6V20M
230x230x100
L6D
D09-03.20
C3
2.28
10
0
5000
2
3000
1.72
-10
40
n
LC
2
0,7
KFP 6V20LM
300x240x100
L6D
D09-03.20
C3
1.86
10
0
5000
2
3000
2.14
-10
40
n
LC
2
0,7
KFP 15V20M
300x240x100
L6D
D09-03.20
C3
2.86
30
0
5000
10
3000
2.14
-10
40
n
LC
2
0,7
KFP 30V20M
400x300x128
L6E
D09-03.21
C3
10.52
50
0
6000
10
3000
4.48
-10
40
n
LC
2
0,7
KFP 60V20M
400x300x128
L6E
D09-03.21
C3
35.52
100
0
6000
20
3000
4.48
-10
40
n
LC
2
0,7
KFP 60V20LM
500x400x137
L6G
D09-03.22
C3
30.98
100
0
6000
20
3000
9.02
-10
40
n
LC
2
0,7
KFP 150V20M
500x400x137
L6G
D09-03.22
C3
90.98
200
0
6000
50
3000
9.02
-10
40
n
LC
2
0,7
KFP 150V20LM
650x500x142
L6G
D09-03.22
C3
136.14
300
0
6000
50
3000
13.86
-10
40
n
LC
2
0,7
KFP 300V20M
650x500x142
L6G
D09-03.22
C3
186.14
500
0
6000
100 3000
13.86
-10
40
n
LC
2
0,7
KFP 600V20AM
800x600x150
PW12B TC5259
C3
340
750
0
6000
100 3000
38
-10
40
n
LC
2
0.7
KFP_V20-BA-e-1419
9