KERN & SOHN KFE-TM User Manual
Page 30

30
KFE-TM-BA_IA-e-1313
10 Instant help
In case of an error in the program process, briefly turn off the display unit and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
Possible cause
The displayed weight does
not glow.
The display unit is not switched on.
Mains power supply interrupted (mains cable
defective).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)