8 non-verifiable models, 9 verifiable models – KERN & SOHN GAB 30K-3PM User Manual
Page 18
6.8 Non-verifiable models:
Switch on the balance and during the selftest press the
and the
button at the same time.
„UnLoad“ is displayed shortly.
Ensure that there are no objects on the weighing pan.
„LoAd“ is displayed
Place adjustment weight, the stability display appears
After the adjustment the balance will carry out a self-test.
„PASS“ will be displayed shortly. Remove adjusting weight
during selftest, the appliance will return into weighing mode
automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.
6.9 Verifiable models:
Switch on balance and press the adjustment switch at the
same time.
After that, during the balance’s selftest, press
and
at the same time.
„UnLoad“ is displayed.
Acknowledge with
. Ensure that there are no objects on
the weighing pan.
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 8)
select the desired setting, the active digit is flashing.
Acknowledge with
. „LoAd“ is displayed.
18
GAB-P-BA-e-1411