8 Troubleshooting


The balance should be switched off for a short time following an interruption in the
programme sequence and disconnected from the mains supply. It is then necessary
to repeat the weighing process from the beginning.


Help:

Interruption Possible

cause

Weight display is not illuminated.

• The balance is not switched on.

• The mains supply connection has been

interrupted (mains cable not plugged
in/faulty).

• Power supply interrupted. .

The weight display changes continu-
ally

• Draught/air movement

• Table/floor vibrations

• The weighing plate is in contact with for-

eign matter.

• Electromagnetic fields / static charging

(choose different location/switch off inter-
fering device if possible)

The weighing result is obviously in-
correct

• The balance display is not set to zero

• Adjustment is no longer correct.

• Great fluctuations in temperature.
• Electromagnetic fields / static charging

(choose different location/switch off inter-
fering device if possible)


Switch the balance off if other error messages should appear and then switch on a-
gain. Contact the manufacturer if the error message does not disappear.