1. Technical Data
• Weighing range:
200 g
•
readability
d:
0.1
g
• operating conditions:
10 – 30 °C
max. 85% air humidity (not condensing)
• power supply:
batteries 2 x LR44
2. Operation
• Turn the cover-disk to position „AUF/OPEN “
• Press the ON/OFF key
; the balance displays “8888” and switches to “0”.
• Place the weighing object in the middle of the weighing platform; the weight in
grams is shown on the display.
• After the weighing process is finished the cover-disk should be turned back to
“ZU/CLOSE”.
• By pressing
the balance can be reset to “0”.
1. Adjustment
• Let loose the ON/OFF-key; the display shows “100”.
• By pressing
for 3 seconds until “CAL” appears on the display.
• Place a 100 g adjusting weight (class M3) on the weighing platform.
• The display switches to “F” and the adjustment is complete.
• If other messages are on the display, the adjustment was faulty. Please repeat
the adjustment process.
2. Important Notes
• Do not overload the balance (when overloaded, error message “E” appears)
• With the occurrence of electromagnetic fields deviations of the displayed
weight value are possible. In case of deviations change location.
• Avoid high air humidity, vapours and dust, because the balance is not
hermetically closed. Do not bring the balance in direct contact with liquids.
Liquids could enter the load cell.
• No permanent load on the weighing platform.
• Never operate the balance in rooms where danger of explosion exists.
• Regularly check the balance with the test weight.
• Warranty expires as soon as the balance is opened.