4.8 Start

up

At start up the instrument performs a lamp test for a few seconds where the display and the

leds blink, in order to verify the integrity and correct operation.

LED

Position Associated

function

Status

out

Relay 1

ON for heater activation

ºC

Setpoint

ON When setting the set point

4.9 Overtemperature

To protect both the unit and your samples, the unit has a cycling thermal switch located inside
the unit and operates at 50

°C. The switch resets when the temperature falls to the reset level

(nominally 15

°C below the operating temperature) .

5 Fault

diagnosis


Symptom Possible

cause Action

required

Unit does not operate

Unit not switched on

Switch on

Unit not plugged into power
supply

Plug in, switch on

Power supply failure

Check that other electrical
appliances on the same
circuit are working

Fuse blown in unit or in plug
(UK units only)

Check and replace - see 7.2


Overtemperature

thermal

switch has operated

Thermal switch operates at
50

°C and resets when the

temperature falls to the reset
level (nominally 15

°C below

the operating temp) .
If the thermal switch keeps
operating, have the unit
checked by a competent
person.

Temperature does not rise
when expected

Set temperature is lower than
block temperature

Check set temperature

Set temperature is too close
to ambient

Raise set temperature

Temperature control circuit
fault

Have unit checked by
competent person

Temperature continues to
rise when not expected

Set temperature is higher
than block temperature

Check setting

Temperature control circuit
fault

Have unit checked by
competent person