beautypg.com

Envirco NSF Biohazard Cabinet User Manual

Page 17

background image

17

Re-certification

It is recommended that the ENVIRCO Biohazard Cabinet be re-certified every six
(6) months and the HEPA filters are replaced if necessary. An Authorized Service
Contractor can provide this service upon request and should be contacted directly.

On-Site Testing Procedure

HEPA Filter Leak Testing

The approved method of testing is to challenge the integrity of the HEPA
filters by introducing a cold polydispersed aerosol

2

upstream from the HEPA

filters while using a light scattering photometer to detect any aerosol
penetration at the downstream surface of the filter.

Test Equipment Required

1. Air operated “Laskin Nozzel” type aerosol generator to produce an

aerosol with a mean particle size of 0.7 micron

2. Light Scattering Aerosol Photometer

3. Silicone RTV

4. Rubber-base solvent cement

5. Silicone stopcock grease

Preparation

1. Remove perforated aluminum diffuser screen at top of work area to

expose the recirculating supply HEPA filter.

2. Open the electrical control panel by removing two bolts at each end. Pull

the rubber tubing off the magnehelic gauge.

3. Disconnect warning buzzer at left-hand side of harness.

4. Remove the work tray.

5. Turn cabinet blower "ON."

6. Locate aerosol generator output hose so that aerosol is fed into the return

air plenum at the back of the work area.

2

NSF 49 (1992 edition)