Airflow
10448 - EBC 47, 4-Ft Type A/B3:
Direct Inflow Method
Type A/B3
-Velocity
100 – 110 fpm
-Airflow
253 – 278 cfm
Calculated Face Velocity
115 – 125 fpm
Exhaust Volume
290 – 316 cfm
Exhaust Filter Area
1.07 sq. ft.
Front Access Area
2.53 sq. ft
Downflow
Velocity
Uniform
72 – 82 fpm
or
Zoned
Zone 1 (Back)
67 – 77 fpm
Zone 2 (Center)
67 – 77 fpm
Zone 3 (Center)
80 – 90 fpm
Zone 4 (Front)
74 – 84 fpm
10276, 6-Ft Type A:
Calculated Face Velocity
Type A
-Velocity
85 – 95 fpm
-Airflow
324 – 362 cfm
Exhaust Filter Area
2.33sq. ft.
Front Access Area
3.81 sq. ft.
Downflow
Velocity
Zone 1 (Back)
54 – 64 fpm
Zone 2 (Center)
53 – 63 fpm
Zone 3 (Front)
67 – 77 fpm
10451, 6-Ft Type B3:
Calculated Face Velocity
Type B3
-Velocity
100 – 110 fpm
-Airflow
381 – 419 cfm
Exhaust Filter Area
2.33sq. ft.
Front Access Area
3.81 sq. ft.
Downflow
Velocity
Zone 1 (Back)
75 – 85 fpm
Zone 2 (Center)
75 – 85 fpm
Zone 3 (Front)
68 –78 fpm