beautypg.com

3 wipe down flows 16 – Envirco BSC User Manual

Page 19

background image

19

Materials Recommended

cleaning

method

Stainless Steel

(a) Wipe the entire surface with a soft paper towel soaked in a concentrated
soap solution such as SWIPE, MR, CLEAN, etc. Then quickly wipe over with
another cloth or towel soaked in clean hot or warm water. The surface will
foam but immediately wipe over again with a dry cloth or towel. The surface
will rapidly dry with a clean, smear-free appearance.
(b)For stained or marked work surface, sinks, etc.,use a mildly abrasive agent
such as SPIC and SPAN. Then wipe clean as above.

Exterior Paint
Finish

Any non-abrasive household agent such as FANTASTIC, BON AMI, etc. Use
a soft cloth or paper towel (non-lining types).DO NOT USE ABRASIVE
AGENTS SUCH AS AJAX OR COMET.

4.2.3 Wipe downflows

First raise the front panel making it convenient to clean the whole interior work area. The type

of wipe-down procedure will depend on all the agents being handled. Consult your Safety

Office and SafetyCommittee for advice.

4.2.4 Base Plenum

Any liquids accidentally released within the work area will either be contained in the work tray

or will flow over the work surface and into the base plenum which surrounds the perimeter of

the work area. This stainless steel base plenum can be drained by means of the ball valve. In

normal use, the drain valve must be in the closed position.

NOTE: Spill capacities are as per NSF-49.

4.2.5 Visual inspection

When moving other equipment, close the safety cabinet and inspect the safety cabinet for any

signs of damage. When using the safety cabinet, the display has readings to record the

pressure. These readings will indicate the loading on the supply HEPA filter.

4.1.6. Regularly challenge the hand intake with a suitable air current tester.