Operation – Boekel Scientific 270500 Thermo Shaker User Manual
Page 6
![background image](/manuals/661622/6/background.png)
SET
STOP 000 000
ACTUAL
HR:MIN:SEC RPM T(C°) TIME RPM T(C°) RUN/STOP
4. Operation
6 7 8 1 2 3 4 5
T(C°)
RPM
4.1
Connect the Boekel Scientific
Groovin’ Tubes
to the AC/DC adapter and the adapter to
the main power supply and the power cord to power supply outlet.
4.2
Insert the tubes into the block.
4.3
Using the TIME buttons
and
(1) set the required operating time (6). If timer
is not set (is set to zero) pressing Run/Stop T(ºC)/RPM button (4) causes continuous
operation of the shaker until the T(ºC)/RPM button is pressed again.
The set time, speed and temperature is shown in the upper line of the display. The actual
time, speed and temperature is shown in the lower line of the display.
4.4
.
Using the RPM buttons
and
(2) set the required shaking speed (7).
4.5
Using the
T(ºC) buttons
a
nd
(3) set the necessary temperature.
The unit then starts heating and maintaining the set temperature as soon as it is
turned on. Decreasing the temperature below 25ºC turns the thermostat OFF (8)
4.6
Press Run/Stop T(ºC)/RPM button (4) to start timer and shaker. The platform motion
starts and the timer indicator will start counting the set time interval (with 1 minute
increments). If the shaking speed is set to zero only the timer will be started.
N2400305, Feb 2007, Rev 0
Page 6
270500
Operating Instructions
TIME
00:28
1250
OFF