6

Heating Rate

Ambient to 70

°C within 25 minutes

Speed of Rotation

6 to 20 RPM


Note: A minimum of at least two tubes are needed for optimum temperature
stability.

8. Maintenance and Service

The Model 230301 & 230301-2 Oven is designed to comply with IEC1010-1.
No routine maintenance is required.

8.1

Cleaning

Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean
up procedures as required for radiation or biohazard control. The outer casing
can be cleaned with a damp cloth. Do not submerge or immerse the Oven in
water.

8.2

Replacement of Fuses

Disconnect the unit from the power supply.
Remove the plug from the socket in the back of the unit.
Pull back on the fuse drawer catch.
Pull out the fuse drawer. (See Figure 2)
Check and replace with the correct fuses if necessary. The fuses should be
5mm x 20mm quick acting, rated 250V.

230301

(115V)

1.0AF

230301-2

(230V)

0.5AF

Push the fuse drawer back in. Reconnect unit to the power supply.

9. Warranty

When used in laboratory conditions and according to these operating instructions
Boekel warrants this product to be free of defective material and workmanship for a
period of two years from the date of manufacture. The liability of Boekel for any
defective equipment during the warranty period shall be limited to the repair of
such equipment or replacement thereof without charge for parts or labor.

10. Service

Should service be required, call Boekel Scientific Customer Service. It is required
to obtain a Returned Material Authorization (RMA) number before any product is
returned for any reason. A Decontamination Certificate must be completed, signed
by the user, and returned to Boekel Scientific prior to receiving the RMA number.
Please be sure to mark the outside of the returned goods package with this RMA
number to ensure prompt handling.

Boekel Scientific

PHONE: (215) 396-8200 or (800) 336-6929

855 Pennsylvania Blvd.

FAX: (215) 396-8264

Feasterville, PA 19053

e-mail: [email protected]