beautypg.com

2 clips for multi-tube carousel (22 clips per bag), Fault diagnosis, Technical specifications – Boekel Scientific 230301-2 Bambino II User Manual

Page 6

background image

5

5.2

Clips for Multi-Tube Carousel (22 clips per bag)

four clips for 50ml tubes
six clips for 15ml tubes
twelve clips for 1.5/2.0ml tubes
Catalog Number 230350

6. Fault Diagnosis

Symptom

Possible Cause

Action Required

1. Unit does not operate

a. Unit not switched on

a. Switch on

b. Unit not plugged into

power supply

b. Plug in, switch on

c. Fuses blown

c. Replace fuses per 8.2

d. Power supply failure

d. Check that other

electrical appliances on
the same circuit are
working

2. Chamber temperature

does not rise when
expected

a. Actual temperature is

higher than set
temperature

a. Check set temperature

b. Temperature control

circuit fault

b. Have unit checked by

competent person

3. Temperature continues

to rise when not
expected

a. Actual temperature is

lower than set
temperature

a. Check set temperature

b. Temperature control

circuit fault

b. Have unit checked by

competent person

4. No rotation, LEDs

illuminated

a. An obstruction prevents

Carousel rotation

a. Remove obstruction

b. Carousel is loose

b. Tighten set screw with

hex key

7. Technical Specifications


This equipment is for indoor use and will meet its performance figures within the
ambient temperature range of 10

°C to 35°C, with maximum relative humidity of 80%

non-condensing. Installation category II (transient voltages). Pollution degree 2 in
accordance with IEC 664. For operation at altitudes of up to 6500 feet.

Temperature range

(Ambient +10

°C) to 80°C.

Setting range

10

°C to 80°C

Stability

+/- 0.1

°C

Overall Accuracy

+/- 0.5

°C at 37°C

Temperature display resolution

0.1

°C

Supply voltage range

Catalog No. 230301 – 115V +/- 10%, 60 Hz.

Catalog No. 230301-2 – 230V +/- 10%, 50/60 Hz.

Power Rating

Model 230301 & 230301-2 90W

This manual is related to the following products: