beautypg.com

Boekel Scientific 234100-2 UV Crosslinker AH User Manual

Page 9

background image

N2400340 Rev. 2

Page 9

7.0

Technical Specifications

This equipment is for indoor use, with maximum relative humidity

of 80% non-condensing. Installation category II (transient

voltages). Pollution degree 2 in accordance with IEC 664. For

operation at altitudes of up to 6500 feet.

Ambient Temperature
Operating Range

16°C to 28°C

Maximum Time Setpoint

999m 59s in 1 second increments

Maximum Energy Setpoint

9,999,900µJ in 100µJ increments

Supply Ratings

Model 234100: 120VAC±10%, 50/60 Hz
Model 234100-2: 230VAC±10%, 50/60 Hz

Power Rating

80W

Net Weight

30lb

Exterior Dimensions (maximum)

15

1

/

8

”W x 14

3

/

8

”D x 10

1

/

4

”H


This product has been tested to the requirements of CAN/CSA-
C22.2 No. 61010-1, second edition, including Amendment 1, or a
later version of the same standard incorporating the same level of
testing requirements.



8.0

Maintenance and Service


8.1

Calibration

This unit has been programmed and calibrated to work with
254nm-, 306nm-, and 368nm-wavelength bulbs, available from
Boekel Scientific. Switching between these wavelengths does
not require manual recalibration of the unit. The proper
wavelength simply needs to be selected from the setup menu.

Note: Using third-party bulbs may compromise the accuracy of
the unit, as the Boekel UV Crosslinker has been programmed and
calibrated to work with Boekel part numbers as listed in the
following table:

254nm Bulb – Clear

Boekel Part Number 920-0305

306nm Bulb – White

Boekel Part Number 920-0306

368nm Bulb – Black

Boekel Part Number 920-0307


This manual is related to the following products: