9.0

Fault Diagnosis / Troubleshooting


Symptom

Possible Cause

Action Required

Unit does not operate

a. Unit not powered on
b. Unit not plugged into

power supply

c. Blown fuse

d. Power supply failure

e. Door is open

a. Press “Power” key
b. Plug in and power on

c. Replace fuse (see

Section 8.5)

d. Check circuit

operation

e. Close door

LCD display is too bright
or too dim

a. Display need contrast

adjustment

a. Refer to Section 8.4,

Contrast Adjustment

Accumulated energy in
Time Mode significantly
lower than usual

a. Blown bulb(s)

a. Ensure all light bulbs

are operational and
replace blown bulbs
as needed

Duration of Energy
Mode significantly
longer than usual for
given amount of energy

a. Blown bulb(s)

a. Ensure all light bulbs

are operational and
replace blown bulbs
as needed